Tissue and blood analysis in patients with orthopedic implant failure: Identification of wear debris and tissue response The hypotheses are:
The hypotheses are:
- The nature of failures of joint prostheses is reflected in blood concentrations of metallic elements.
- Degradation products in periprosthetic tissue from failed joint prostheses provide information about the failure mechanism.
Recent studies have shown that large amounts of minute nano- to micron-sized wear particles are produced by orthopedic implants, setting into motion a cascade of events that ultimately may result in osteolysis, possibly leading to failure of artificial joints. Determination of the concentrations of metal particles and ions in tissue and blood may be useful in the diagnosis of patients with symptoms after total joint replacement, as increased levels can be indicative of mechanical dysfunction or wear (Høl et al. 2006; Høl et al. 2008).
Tissue surrounding the joint along with blood samples are collected and stored in the established Peri-prosthetic Tissue Bank. The degradation products in the biological material are characterized and quantified using:
- Optical microscopy
- Scanning- and transmission EM
- ICP-MS trace element analysis
Additionally, retrieved tissue will be analyzed for biological reactions, such as lymphocyte and macrophage responses, by immunohistochemical techniques. The cellular effects of nanoparticles released from surgical implants are currently being studied in a parallel set of studies related to nanotoxicology (external link), an activity that is associated with our group.
Completed project:
Immune response to implants
It is controversial whether implanted material can cause clinical immune-mediated reactions in patients. The substances involved are largely unknown, and not necessarily covered by the standard dermatological patch tests commonly employed for assessment of skin sensitivity. Based on metal-based test substances developed by us (Vamanu et al. 2008), experimental patient testing using flow cytometry to assess lymphocyte stimulation and multiplex bead assay to assess monocytic and lymphatic cytokines and chemokines has been explored (Høl PJ, Kristoffersen EK, Gjerdet NR, Pellowe AS. Novel Nanoparticulate and Ionic Titanium Antigens for Hypersensitivity Testing. Int J Mol Sci. 2018;19(4). doi: 10.3390/ijms19041101).
Participants:
Paul J. Høl (UiB/Helse Bergen); Geir Hallan (UiB/Helse Bergen); Kari Indrekvam (UiB/Helse Bergen-Kysthospitalet).